Lot Number 0027072300 |
Device Problems
Failure to Advance (2524); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by manufacturer: synergy xd mr ous 3.00 x 20mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Stent struts on the distal end were deformed and damaged.The undamaged stent od (outer diameter) was measured within max crimped stent profile measurement specification.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2021.It was reported that crossing difficulties were encountered.A 3.00 x 20mm synergy xd balloon expandable stent was selected to treat a lesion in the left anterior descending artery.The synergy xd could not cross the lesion due to the calcification.The procedure was completed with another same device.No patient complications were reported.However, returned device analysis revealed stent damage.
|
|
Search Alerts/Recalls
|