Device evaluated by mfr: a synergy xd mr ous 3.50 x 32mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage with the distal stent struts lifted.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent od (outer diameter) was measured and the results were within maximum crimped stent profile measurement.The balloon body was reviewed; no issues were noted on the balloon cones.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the tip, hypotube, and shaft showed no signs of damage.
|