Model Number 482317550
Medical Device Problem Codes Fracture (1260); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the tulip of a xia 3 polyaxial screw disengaged from the screw shaft approximately 5 years post-operatively.Revision surgery was performed.
 
Additional Manufacturer Narrative
Visual inspection was performed which confirmed screw shank fractured just below the shank bulb.Device and complaint history records were reviewed for this lot, no relevant manufacturing issues or similar complaints were identified.The device stg was reviewed and the following was found to be relevant: the surgeon must warn the patient of the surgical risks and make aware of possible adverse effects.The surgeon must warn the patient that the devices cannot and do not replicate the flexibility, strength, reliability or durability of normal healthy bone, that the implants can break or become damaged as a result of strenuous activity or trauma, and that the devices may need to be replaced in the future.If the patient is involved in an occupation or activity which applies inordinate stress upon the implant (e.G., substantial walking, running, lifting, or muscle strain) the surgeon must advise the patient that resultant forces can cause failure of the devices.Once implanted, the implants are subjected to stresses and strains.These repeated stresses on the implants should be taken into consideration by the surgeon at the time of the choice of the implant, during implantation as well as in the post-operative follow-up period.Indeed, the stresses and strains on the implants may cause metal fatigue or fracture or deformation of the implants, before the bone graft has become completely consolidated.This may result in further side effects or necessitate the early removal of the osteosynthesis device.The most likely cause of the reported event was determined to be the age of implantation and normal wear.Device was active and served its intended purpose for 5 years.Due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone.Excessive strenuous activity or trauma can contribute to screw fractures, but was not confirmed.
 
Event or Problem Description
It was reported that the tulip of a xia 3 polyaxial screw disengaged from the screw shaft approximately 5 years post-operatively.Revision surgery was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM POLYAXIAL SCREW DIA 7.5 X 50 MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13026848
Report Number0009617544-2021-00238
Device Sequence Number17503284
Product Code NKB
UDI-Device Identifier04546540562364
UDI-Public04546540562364
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K142381
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number482317550
Device Catalogue Number482317550
Device Lot NumberB56285
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date12/17/2021
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;