Catalog Number UNK_STM
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies. There have been no other similar events for the lot referenced. Device not returned.
 
Event Description
As reported: "i had a problem with the extension of the prosthesis pin (b)(4) (left distal femoral jts). It started with the regular sound of gear movement but after 30s changed to a humming sound. I fallowed the trouble shooting protocol but it did not help. We tried repositioning the leg, turning to b that gave gear movement sound and then it worked for several seconds and again started humming, we let her walk. " update (b)(6)2021 wg: additional information received from distributor: i had another extension. And there the prosthesis extended the 4mm we intended, so, i do not think the problem is with the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISTAL FEMUR JTS
Type of DevicePROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK WD6 3SJ
Manufacturer (Section G)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK WD6 3SJ
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key13029354
MDR Text Key285481322
Report Number3004105610-2021-00162
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K133152
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation
Type of Report Initial
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/31/2020
Device Catalogue NumberUNK_STM
Device Lot NumberPIN 22139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/17/2021 Patient Sequence Number: 1
-
-