| Model Number |
626-00-46F |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
11/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a post-op x-ray taken the same day as the primary hip replacement indicated dislocation.A closed reduction was attempted and failed, and the patient was revised the same day.During the revision, the adm/ mdm poly insert was found floating in the joint.The 46f adm/ mdm poly insert and 28+0 biolox head were revised to another 46f adm/ mdm insert and 28 +4 biolox head.Intraoperatively, the biolox head was still attached to the stem trunnion.
|
| |
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
|
| |
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
|
| |
|
Event or Problem Description
|
|
It was reported that a post-op x-ray taken the same day as the primary hip replacement indicated dislocation.A closed reduction was attempted and failed, and the patient was revised the same day.During the revision, the adm/ mdm poly insert was found floating in the joint.The 46f adm/ mdm poly insert and 28+0 biolox head were revised to another 46f adm/ mdm insert and 28 +4 biolox head.Intraoperatively, the biolox head was still attached to the stem trunnion.
|
| |
|