Model Number 022-275
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Paresis (1998); Presyncope (4410)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Similar complaints for this issue were trended for the test strip lot.It was concluded that the number of complaints for the lot did not breach thresholds indicative of a systemic issue.In addition, similar complaints for this issue were trended including the reported meter.It was concluded that the number of complaints for the meter did not breach thresholds indicative of a systemic issue.
 
Event or Problem Description
On (b)(6) 2021, the patient/lay user contacted lifescan (lfs) usa alleging that his onetouch verio iq meter read inaccurately high compared to his feelings and/or normal results.The complaint was classified based on the customer care agent (cca) documentation.The patient reported that the alleged issue began on (b)(6) 2021, in the evening.The patient claimed obtaining a blood glucose reading of ¿80 mg/dl¿ with the subject meter and indicated that after the reading he ¿had a faint, could not move¿ and had to call for help.Within 30 minutes the paramedics arrived and they obtained a blood glucose reading of ¿43 mg/dl¿ on an emergency medical services (ems) meter.The patient manages his diabetes with diet and exercise and claimed that he did not make any changes to his usual diabetes management regimen in response to the alleged issue.The patient informed the cca that he was treated by an hcp with a glucose injection.During troubleshooting, the cca confirmed that the unit of measure was set correctly on the subject meter, the patient had used an approved sample site to obtain the blood samples and that the patient was following the correct testing procedure.The cca noted that the patient did not have control solution at the time of the call to test the subject system.The cca established that the test strip vial was intact, that the test strips had been stored properly, were not open beyond their discard date and had not expired.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed symptoms suggestive of a serious injury adverse event and because the patient reportedly received hcp treatment for an acute low blood glucose excursion after the alleged product issue began.The subject meter could not be ruled out as a cause or contributor to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO IQ METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN SCOTLAND
beechwood park north
inverness IV2 3 ED
UK   IV2 3ED
Manufacturer Contact
simon palmer
beechwood park north
inverness IV2 3-ED
UK   IV2 3ED
1463383679
MDR Report Key13031081
Report Number3008382007-2021-04281
Device Sequence Number14595525
Product Code NBW
UDI-Device Identifier00353885008402
UDI-Public00353885008402
Combination Product (Y/N)N
PMA/510(K) Number
K110637
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number022-275
Device Catalogue Number022-275
Device Lot Number4762937
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2021
Initial Report FDA Received Date12/17/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age66 YR
Patient SexUnknown
-
-