| Model Number |
OPT970 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor in (b)(4) reported via a fisher and paykel (f&p) field representative that the tubing of an opt970 optiflow + tracheostomy direct connection was found damaged during patient use.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in canada reported via a fisher and paykel (f&p) field representative that the tubing of an opt970 optiflow + tracheostomy direct connection was found damaged during patient use.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The opt970 optiflow + tracheostomy direct connection interface is used to deliver humidified oxygen to patients via tracheostomy.The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's trache.Method: the complaint opt970 optiflow + tracheostomy direct connection interface was returned to fisher & paykel healthcare (f&p) new zealand for investigation, where it was visually inspected.Results: visual inspection of the returned cannula revealed that the tubing was stretched and pulled apart close to the manifold.The customer reported that the complaint cannula's lanyard was not used.Conclusion: we are unable to determine the cause of the reported damage to the subject opt970 optiflow + tracheostomy direct connection interface.However the damage observed was likely caused by the tubing being pulled.The customer reported that the complaint cannula's lanyard was not used which might have contributed to the reported event.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject cannula would have met the specification at the time of production.The user instructions which accompany the opt970 optiflow + tracheostomy direct connection interface show in pictorial format the correct placement and fitting of the cannula and also warn: "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death."."do not crush or stretch tube, to prevent loss of therapy."."failure to use the set-up described above can compromise performance and affect patient safety.".
|
| |
|