| Model Number |
BOET3213 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
Doctor reported the implant was removed during the implant placement because when he was trying to screw with the ratchet, it fracture the hex of the implant.It was not possible to remove the implant even screwing backwards, that is why the threads end very damaged.Doctor did not place other implant in the same surgery.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The following sections have been updated: h3: device evaluated by manufacturer: change ¿no' to 'yes'.An implant was returned for investigation, in the complaint.Visual inspection of the as returned product identified worn markings due to usage and a fracture at the collar.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.No pre-existing conditions were noted on the per.The reported device location and length of usage is unknown.Pictures or x-ray images were not provided.Review of appropriate documentation: documents reviewed: biomet 3i dental implant ifu (p-iis086gi) rev h - july 2021 information identified: 'warnings' 'precautions' 'potential adverse events'.Dhr review: dhr review was completed for the subject lot number (2016120112).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2016120112) for similar event (complaint category keyword: dental : functional : fracture : implant) and no other complaint was identified.Review completed utilizing keywords: fracture implant.Post market trend review: april post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
|
| |
|
Search Alerts/Recalls
|