DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ3 RMLL; EARLY INTERVENTION : KNEE TIBIAL TRAY Back to Search Results
Model Number 1024-52-300
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fatigue (1849); Pain (1994); Loss of Range of Motion (2032)
Date of Event 10/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was revised to treat sudden sharp pain and giving way of the knee secondary to insert disassociation.Doi: (b)(6) 2012; dor: (b)(6) 2021 unk knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and evidence of implant disassociation was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mrae) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB TRAY SZ3 RMLL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13042784
Report Number1818910-2021-28393
Device Sequence Number5424323
Product Code HRY
UDI-Device Identifier10603295002390
UDI-Public10603295002390
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K070267
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1024-52-300
Device Catalogue Number102452300
Device Lot Number17946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer12/14/2021
01/21/2022
Initial Report FDA Received Date12/20/2021
Supplement Report FDA Received Date01/06/2022
01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIGMA HP UNI FEMORAL SZ3 RM/LL.; SIGMA HP UNI INS SZ3 7MM RM/LL.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexUnknown
-
-