• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE Back to Search Results
Model Number 8900-4006
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to cardiovert a patient (age & gender unknown), a spark was emitted from the electrode pad.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The propadz radiolucent electrodes were returned to zoll medical corporation for evaluation.Inspection of the pads observed evidence that a spark occurred due to an air pocket between the pads and the patient.The electrodes were scrapped at zoll.Electrode labeling states the importance of good placement on the patient and provides instruction for proper electrode application technique.Poor adherence and/or air under the electrodes can lead to the possibility of sparking/arcing and skin burns.The instructions for use on the pro-padz radiolucent electrodes label states, "apply electrode by attaching one edge securely to the patient and rolling smoothly from that edge to the other.Be careful not to trap any pockets of air between the gel and skin." analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE
Type of Device
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key13043026
MDR Text Key282714985
Report Number1218058-2021-00144
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946016661
UDI-Public00847946016661
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/05/2023
Device Model Number8900-4006
Device Catalogue Number8900-4006
Device Lot Number3221A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-