Model Number PL572T
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was an issue with pl572t-ligature clip 12 mag.= 144 pcs.According to the complaint description, it was reported that it was able to clip a few shots without any problems, but the clip (cartridge) came off in the middle.There was no described patient harm.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Additional information/correction: b5- update based upon investigation results, this event was re-evaluated and is considered no longer reportable due to risk file- no malfunction or serious injury.Investigation results: visual investigation: we made a visual inspection of the product; here we detected deformed latches of the slider sheet.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 1(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Explanation and rationale: if the cartridge is engaged not completely or was damaged during inserting, there is an impairment of product functionality.This could lead to deformed latches of the slider sheet, to clip jam and can cause the clip-cartridge to jump off.Based upon historically grown product experience and due to different simulation regarding a cartridge which is engaged not completely, this leads to the described errors.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigation results, a capa is not required.
 
Event or Problem Description
Update: the malfunction occurred in the department of urology, during an unspecified procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13043137
Report Number9610612-2021-00779
Device Sequence Number14805010
Product Code FZP
UDI-Device Identifier04038653341856
UDI-Public4038653341856
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2025
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52614245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2021
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date12/20/2021
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-