• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. PLUG; EBI 5.5 HELICAL FLANGE SPINAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. PLUG; EBI 5.5 HELICAL FLANGE SPINAL SYSTEM Back to Search Results
Catalog Number 2000-1005
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2021-00467.
 
Event or Problem Description
It was reported five polaris plugs stripped intraoperatively.They were removed and replaced without patient impacts.This is report one of five for this event.
 
Event or Problem Description
It was reported five polaris plugs stripped intraoperatively.They were removed and replaced without patient impacts.This is report one of five for this event.
 
Additional Manufacturer Narrative
Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned devices match the information in the complaint file and were examined.There were five returned closure tops, however one was lost in the wash process (lot j6427145) and is unable to be inspected.The four that were able to be inspected revealed stripped hex on three (lot numbers j6364237, j6392969 and j6435242) and one with damaged threads (lot j6364237).Potential cause: root cause was unable to be determined.This event could possibly be attributed to applying off-axis forces during insertion.Dhr review: per dhr review, the parts were likely conforming when they left zimmer biomet control.Device use: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG
Common Device Name
EBI 5.5 HELICAL FLANGE SPINAL SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key13045812
Report Number3012447612-2021-00463
Device Sequence Number5364683
Product Code MNH
UDI-Device Identifier00888480207115
UDI-Public(01)00888480207115(11)181127(10)J6427145
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K061441
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2000-1005
Device Lot NumberJ6392969
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date12/20/2021
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight80 KG
-
-