| Catalog Number |
2000-2450 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported a polyaxial screw lot its polyaxial motion intra-operatively.It was removed and replaced with another screw to complete the procedure without patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed surface scratches in the anodize of the returned screw (2000-2450), but no other signs of damage.A functional inspection revealed the tulip has no polyaxial motion.Potential cause root cause was unable to be determined.This event could possibly be attributed to applying off-axis forces during insertion.Dhr review per dhr review, the parts were likely conforming when they left zimmer biomet control.Device use this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
| |
|
Event or Problem Description
|
|
It was reported a polyaxial screw lot its polyaxial motion intra-operatively.It was removed and replaced with another screw to complete the procedure without patient impacts.
|
| |
|