• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION AVIATOR PLUS .014 5.0X20 142CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION AVIATOR PLUS .014 5.0X20 142CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4245020W
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the balloon of an aviator plus.014 5.0x20 142cm percutaneous transluminal angioplasty (pta) balloon catheter burst after it was inflated several times within the specified pressure.It was replaced with a new unknown aviator and the procedure was completed.There was no reported patient injury.The procedure was a "bpa" case.The device will not be returned for evaluation because it was discarded.
 
Additional Manufacturer Narrative
The product history review is expected but has not been completed.Additional information pending and will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: complaint conclusion: as reported, the balloon of an aviator plus.014 5.0x20 142cm percutaneous transluminal angioplasty (pta) balloon catheter burst after it was inflated several times within the specified pressure.It was replaced with a new unknown aviator and the procedure was completed.There was no reported patient injury.The procedure was a "bpa" case.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82228500 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics and procedural factors, although unknown, may have contributed to the reported event.However, without the return of the device for analysis and with the limited amount of information provided, it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to use, the device should be examined to verify functionality and integrity and ensure that its size is suitable for the specific procedure.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in-vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure-monitoring device is recommended to prevent over-pressurization.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon¿.Neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR PLUS .014 5.0X20 142CM
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key13046319
Report Number9616099-2021-05210
Device Sequence Number5449622
Product Code LIT
UDI-Device Identifier20705032005828
UDI-Public20705032005828
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K071189
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model Number4245020W
Device Catalogue Number4245020W
Device Lot Number82228500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/24/2021
Supplement Date Received by Manufacturer12/20/2021
Initial Report FDA Received Date12/20/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN AVIATOR
Patient SexUnknown
-
-