| Model Number |
5444-0-310 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Cerebral Edema (4403); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.Device not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had most anterior fixation as well as most posterior fixation pulling out from patient¿s bone.The surgeon stated that this was possibly due to pressure from brain swelling.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint event could be confirmed.The complaint department has already covered two previous reported events with a medpor customized product (lot# 2007161015) for this patient.According to the previous investigations, the patient had back-to-back issues with brain swelling.The surgeon used an external ventricular drain (evd) to drain extra cerebrospinal fluid (csf) from the brain.Even though the implant was fixated appropriately with more screws and plates than recommended, the implant loosened due to the pressure of the brain fluid in the area where the swelling was present.Furthermore, a review of the manufacturing records confirmed the lot meets all final release specifications.Summarizing all obtained information, the root cause of the reported event was the swelling present laterally and posteriorly to the patients left eye prior to the implant.Additionally, it is the surgeon¿s responsibility to reduce the swelling prior to implanting the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had most anterior fixation as well as most posterior fixation pulling out from patient¿s bone.The surgeon stated that this was possibly due to pressure from brain swelling.
|
| |
|