| Model Number |
60005 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
08/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Site reported that a light adjustable lens was explanted from the patient as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's first awareness of the event was 12/1/2021.Device history record for the lens was reviewed.No issues were noted.The lens is in the process of being returned for evaluation.
|
| |
|
Event or Problem Description
|
|
Site reported that a light adjustable lens was explanted from the patient as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's first awareness of the event was on 12/1/2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Site reported that a light adjustable lens was explanted from the patient as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's first awareness of the event was (b)(6) 2021.The explanted lens was returned for evaluation.Visual inspection and optical testing of the returned lens confirmed the presence of an ambient zone on the lens, which is indicative of exposure of the lens to ambient uv light prior to lock in.
|
| |
|
Event or Problem Description
|
|
Site reported that a light adjustable lens was explanted from the patient as the desired refractive outcome was not achieved after light adjustment treatments.Rxsight's first awareness of the event was (b)(6) 2021.
|
| |
|