• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DHS®/DCS® GUIDE SHAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DHS®/DCS® GUIDE SHAFT Back to Search Results
Model Number 338.21
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: event year is reported as 2021; however exact date of event is unknown.Complainant part is expected to be returned for manufacturer review/ investigation but has yet to be received.Reporter is a j&j representative.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on unknown, the insertion screwdriver pieces are bent causing them to not properly fit in the insertion handle.It is unknown when and where the issue was discovered.It is unknown if there is patient nor procedure involvement.This complaint involves four (4) devices.This report is for (1) ddhs®/dcs® guide shaft.This report is 4 of 4 for (b)(4).
 
Event or Problem Description
It was reported that on unknown, the insertion screwdriver pieces are bent causing them to not properly fit in the insertion handle.It is unknown when and where the issue was discovered.It is unknown if there is patient nor procedure involvement.This complaint involves four (4) devices.This report is for (1) ddhs®/dcs® guide shaft.This report is 4 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Event year is reported as 2021; however exact date of event is unknown.Complainant part is expected to be returned for manufacturer review/ investigation but has yet to be received.Reporter is a j&j representative.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Date device returned to manufacturer.Codes updated to imdrf codes.Visual inspection: the complaint device (dhs/dcs guide shaft) was returned for investigation.It was that the device is bent.Assembling issues can be attributable to this condition, therefore, the complaint is confirmed.Additionally, scratches were found along the device.No other issues were identified.Functional test: a complete functional test could not be performed as the device was returned by itself without mating part.Dimensional inspection: due to lot number being unknown a dimensional inspection cannot be performed to its manufactured revision.Investigation conclusion: the reported condition of the complaint device (dhs/dcs guide shaft) is confirmed as the shaft is bent.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/ specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot - a manufacturing record evaluation cannot be performed due to lot number being unknown.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DHS®/DCS® GUIDE SHAFT
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13048724
Report Number2939274-2021-07176
Device Sequence Number11750194
Product Code FZX
UDI-Device Identifier10886982193247
UDI-Public(01)10886982193247
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number338.21
Device Catalogue Number338.21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/22/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date12/20/2021
Supplement Report FDA Received Date01/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DHS/DCS COUPLING SCREW; DHS/DCS COUPLING SCREW; DHS/DCS GUIDE SHAFT
Patient SexUnknown
-
-