• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Gas/Air Leak (2946); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) that a rt380 adult dual-heated evaqua2 breathing circuit failed the ventilator leak test before use.There was no patient involvement.
 
Manufacturer Narrative
Ps(b)(4).Method: the complaint rt380 adult dual-heated evaqua2 breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected and pressure tested.Results: visual inspection of the complaint rt380 adult dual-heated evaqua2 breathing circuit revealed no damage to the returned breathing circuit or the dryline.The pressure test also revealed that the breathing circuit was within specification.Conclusion: we were unable to determine what may have caused the failed leak test as reported by the customer, as no fault was found with the returned device.The reported leak most likely occurred due to a loose connection in the setup or as the water feedset of the mr290 vented autofeed humidification chamber was still in the feedset winder and not connected to a water bag when the leak test was performed.The feedset winder is designed to secure the feedset and spike during transport and storage.It is not designed to seal the spike.All rt380 adult dual-heated evaqua2 breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject infant breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit show in pictorial format the correct set-up of the circuit and also states the following: - "check all connections are tight before use." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that a rt380 adult dual-heated evaqua2 breathing circuit failed the ventilator leak test before use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13050281
MDR Text Key284455128
Report Number9611451-2021-01439
Device Sequence Number1
Product Code BZE
UDI-Device Identifier09420012429728
UDI-Public(01)09420012429728(10)2101561596(11)210319
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot Number2101561596
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received03/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P MR850 RESPIRATORY HUMIDIFIER.; F&P MR850 RESPIRATORY HUMIDIFIER.
-
-