Catalog Number 8000.COM05 |
Device Problems
Smoking (1585); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The complaint is under investigation.
|
|
Event Description
|
It was reported that at the time of replacing the pump module, while turning the pump on, a burning smell is observed coming from the pump.No actual patient harm or operator harm occurred.
|
|
Manufacturer Narrative
|
In regard to this event an eva pump module was returned for investigation.Investigation of the returned module revealed a burn mark on the power converter of the mainboard of the pump module.Historical review of the complaint database indicated that no similar complaints have been logged on the eva surgical system subject to this complaint.Based on the investigation findings it was determined that the issue occurred due to a component failure of the power converter on the mainboard of the pump module.Trend analysis indicates that the product is performing within anticipated rates.Complaints will be closely monitored to identify any significant adverse trends.The risk identified is included in the risk management documentation.Trend analysis indicates that the product is performing within anticipated rates.Therefore, no remedial or corrective/preventive actions will be undertaken at this moment.Complaints will be closely monitored to identify any significant adverse trends.All similar incidents related to the eva surgical system are included in the analysis.Since 2018 more than 750.000 surgeries have been performed with the eva surgical systems installed.
|
|
Event Description
|
It was reported that at the time of replacing the pump module, while turning the pump on, a burning smell is observed coming from the pump.No actual patient harm or operator harm occurred.
|
|
Search Alerts/Recalls
|