C.R. BARD, INC. (COVINGTON) -1018233 SILICONE FLAT DRAIN Back to Search Results
Model Number 0070370
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the holes in the front and back of the flat drain tube were misaligned.As per follow-up information received on 07-dec-2021, no other damage was found.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The product had caused the reported failure.Visual evaluation of the returned sample noted one unopened (with original packaging), hubless drain.A potential root cause for this failure could be ¿incorrect set-up".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications for use: wound drains are used to remove exudates from wound sites.Contraindications: do not use for chest drainage.Warning: do not bypass/inactivate anti-reflux valve." the actual/suspected device was evaluated.
 
Event or Problem Description
It was reported that the holes in the front and back of the flat drain tube were misaligned.As per follow-up information received on 07dec2021, no other damage was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE FLAT DRAIN
Common Device Name
SILICONE FLAT DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13051519
Report Number1018233-2021-08250
Device Sequence Number3888110
Product Code GBX
UDI-Device Identifier00801741090745
UDI-Public(01)00801741090745
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model Number0070370
Device Catalogue Number0070370
Device Lot NumberNGFS4564
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2021
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date06/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-