Model Number 406136
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2021
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, a blood leak was noted from the hemostasis valve.The sheath was replaced and the procedure was completed with no adverse patient consequences.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3.One 14f fast-cath introducer sheath was received for evaluation.Functional testing confirmed a fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing, resulting in a hole, was noted in at both ends of the distal seal slit.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak remains unknown.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH HEMOSTASIS INTRODUCER
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13051561
MDR Text Key282895017
Report Number3005334138-2021-00779
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734014137
UDI-Public05414734014137
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K914090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model Number406136
Device Catalogue Number406136
Device Lot Number7591193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight70 KG
-
-