RESPIRONICS, INC. TRILOGY O2; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1040004
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for service.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, a "service required" code was found in the ventilator's downloaded error log.The device's oxygen blending module board was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY O2
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge drive
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge drive
murrysville PA 15668
Manufacturer Contact
kimberly shelly
head of quality, src
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13051575
Report Number2518422-2021-08255
Device Sequence Number11738142
Product Code CBK
UDI-Device Identifier00606959033887
UDI-Public00606959033887
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1040004
Device Catalogue Number1040004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2021
Initial Report FDA Received Date12/21/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-