• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915189
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2021
Event Type  malfunction  
Event Description
The pivot pin got loose and the tip of the jaws got damaged during pretest before use.
 
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
The pivot pin got loose and the tip of the jaws got damaged during pretest before use.
 
Manufacturer Narrative
(b)(4) the dhr for the returned device was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha wi facility as part of a 50-piece lot in april of 2020.Evaluation of the returned instrument shows that the tips are loose and misaligned, and the jaw pivot pin is pulled thru one side of the damaged/bent outer tube assembly.We are able to validate this complaint.We are unable to determine what caused the jaws to be loose and misaligned and for the jaw pivot pin to be pulled thru one side of the bent/damaged outer tube assembly but mishandling of this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key13052620
MDR Text Key282563321
Report Number3011137372-2021-00350
Device Sequence Number1
Product Code GDO
UDI-Device Identifier24026704696813
UDI-Public24026704696813
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN915189
Device Catalogue Number544965
Device Lot Number06G1990518
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2021
Initial Date FDA Received12/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-