• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Model Number 1011344-40
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during preparation of the 7-10x40mm rx.014 acculink self-expanding stent system (sess), the device was being flushed and liquid sprayed out the side of the device (hub).Another same size acculink was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported leak was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that inadvertent mishandling during repackaging and/or during preparation for use resulted in the noted gap between handle halves; thus resulting in the reported leak; however this cannot be confirmed.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.The noted multiple bends and kinks on the entire length of the shaft likely occurred due to handling or during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13052732
Report Number2024168-2021-11952
Device Sequence Number13855282
Product Code NIM
UDI-Device Identifier08717648076398
UDI-Public08717648076398
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number1011344-40
Device Catalogue Number1011344-40
Device Lot Number1090861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/02/2021
Supplement Date Received by Manufacturer01/06/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-