|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 21, 2021.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date and lot number).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H4 (device manufacture date).H6 (identification of evaluation codes 3331, 4114, 11, 3259, 4307).Type of investigation: #1: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #2: 3331 - analysis of production records.Type of investigation #3: 4114 - device not returned.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The affected sample was not returned; however, a picture was provided that showed the arterial thermistor being broken off.A representative retention sample was reviewed and found to have no damage to the oxygenator, specifically in the area of the arterial thermistor.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|