EXACTECH, INC. EQUINOXE; GLENOID IMPACTOR TIP Back to Search Results
Catalog Number 315-07-06
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported, during the surgeon¿s routine impaction of an anatomic glenoid implant, the impactor tip broke into two pieces which were both easily retrieved.Nothing fell into the wound site.The (b)(6) male was not affected by the break.Patient was last known to be in stable condition following the event.Device is to be returned.
 
Additional Manufacturer Narrative
Pending evaluation.
 
Additional Manufacturer Narrative
Section h10: (d4) lot number: 92919002 (h3) based on previous complaint investigations, the reported event, and returned device, the fractured glenoid impactor tip reported in was likely the result of crack initiation, propagation, and ultimate fracture of the device which was likely caused by abnormal or aggressive use inconsistent with the intended use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
GLENOID IMPACTOR TIP
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key13052787
Report Number1038671-2021-00715
Device Sequence Number3888142
Product Code KWT
UDI-Device Identifier10885862251824
UDI-Public10885862251824
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number315-07-06
Device Lot Number92919002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/30/2021
Supplement Date Received by Manufacturer01/06/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date01/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
-
-