RESPIRONICS INC. Sensor; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1134711
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilator was alarming for a faulty flow sensor.There was no harm or injury reported.The device's external flow sensor was replaced to address the issue.The flow sensor was returned to the manufacturer for evaluation.The external flow sensor was found to have rust on the internal mesh.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensor
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
MDR Report Key13053349
Report Number2518422-2021-08357
Device Sequence Number1662410
Product Code CBK
UDI-Device Identifier00606959050839
UDI-Public00606959050839
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1134711
Device Catalogue Number1134711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2021
Initial Report FDA Received Date12/21/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-