| Model Number |
W5-4-2-MVI-3 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device remains impanted in the patient; therefore, the alleged product issue cannot be confirmed.The instructions for use (ifu) identifies migration or misplacement as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the web device was used for treatment of a pre-ruptured acomm aneurysm.The aneurysm re-ruptured after web deployment.An angiogram run showed the aca was shut down and the web was protruding into the left a1.At that point bp was 144/37 from 120/55 when stent first detached.Bp pumped up with 2 boluses of iv integrilin and heparin 10,000 units.The physician attempt to aspirate with a catheter without success and a lvis jr.Stent was placed with good opposition against the wall to completed the procedure.The patient current condition described as "patient is good".
|
| |
|