Model Number W5-4-2-MVI-3
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device remains impanted in the patient; therefore, the alleged product issue cannot be confirmed.The instructions for use (ifu) identifies migration or misplacement as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that the web device was used for treatment of a pre-ruptured acomm aneurysm.The aneurysm re-ruptured after web deployment.An angiogram run showed the aca was shut down and the web was protruding into the left a1.At that point bp was 144/37 from 120/55 when stent first detached.Bp pumped up with 2 boluses of iv integrilin and heparin 10,000 units.The physician attempt to aspirate with a catheter without success and a lvis jr.Stent was placed with good opposition against the wall to completed the procedure.The patient current condition described as "patient is good".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Common Device Name
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key13054750
Report Number2032493-2021-00512
Device Sequence Number14425010
Product Code OPR
UDI-Device Identifier00842429110812
UDI-Public(01)00842429110812(11)210917(17)240831(10)0000085474
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P170032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberW5-4-2-MVI-3
Device Catalogue NumberW5-4-2
Device Lot Number0000085474
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2021
Initial Report FDA Received Date12/21/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
Patient SexMale
Patient Weight98 KG
-
-