SMITH & NEPHEW ORTHOPAEDICS AG PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Model Number 75004613
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that, during total hip replacement, a plus sl-plus mia dbl offset adapt.Right 60/25mm broke on impaction, inside the patient.All parts were retrieved by hand.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during total hip replacement surgery, a plus sl-plus mia dbl offset adapt.Right 60/25mm broke on impaction, inside the patient.The device, used in treatment, was returned for investigation.Upon visual inspection, the reported failure mode could be confirmed.The clutch is broken.Apart from that, the device shows normal signs of usage such as gouges and dents.A complaint history review was performed.One additional complaint with the batch in scope was reported so far.A review of the batch record revealed no deviations from the standard manufacturing process, that could have contributed to the reported issue.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.Upon performed medical investigation, no patient injuries or adverse consequences were reported.Therefore, no further clinical/medical assessment is warranted at this time.The failure mode and the severity are covered in the corresponding risk management document.A review of past corrective actions was performed.No further escalation is required.However, based on the performed investigations, the relationship between the reported event and the device was confirmed.This instrument is designed to hold and guide the detachable rasps for broaching.It is therefore subjected to repeated impact forces via the modular knock plate or the pneumatic woodpecker.Previous investigations demonstrated, that under specific circumstances, this device may fracture during impaction.An optimized design of the device has been released in order to reduce the occurrence of this issue.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all instruments must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13054835
Report Number9613369-2021-00425
Device Sequence Number3951370
Product Code KWY
UDI-Device Identifier07611996085471
UDI-Public07611996085471
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75004613
Device Catalogue Number75004613
Device Lot NumberB51295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2021
Supplement Date Received by Manufacturer01/20/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-