• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 15MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 15MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 11/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that 270 days post implant with the subject flow diverter in a clinical trial procedure to treat an unruptured saccular intracranial aneurysm measuring 4.5mm in the right internal carotid artery-ophthalmic (c6 segment), the patient had in stent stenosis that had increased from 0-25% post implantation to 50-75% at 6 months follow up.During the procedure, the subject flow diverter was reported to have been opened and used successfully to completely cover the aneurysm neck.No treatment was administered for the reported event.The patient did not experience any clinical complications since the last visit.No other information is available.
 
Additional Manufacturer Narrative
Device is implanted in patient.
 
Event or Problem Description
It was reported that 270 days post implant with the subject flow diverter in a clinical trial procedure to treat an unruptured saccular intracranial aneurysm measuring 4.5mm in the right internal carotid artery-ophthalmic (c6 segment), the patient had in stent stenosis that had increased from 0-25% post implantation to 50-75% at 6 months follow up.During the procedure, the subject flow diverter was reported to have been opened and used successfully to completely cover the aneurysm neck.No treatment was administered for the reported event.The patient did not experience any clinical complications since the last visit.No other information is available.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported that 6 months after a clinical trial procedure with the subject flow diverter, the patient had post implantation in-stent stenosis that increased to 50-75%.Additional information provided by the customer indicated that the device was successfully implanted without any device deficiencies, the post implantation in-stent stenosis was 0-25%, the subject did not experience any new adverse events or changes to existing adverse events since the last visit, unruptured saccular intracranial aneurysm measuring 4.5mm was being treated with the subject device, and the event occurred in the right internal carotid artery-ophthalmic (c6 segment).Based upon medical review, the harm observed in this complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted within the dfu, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 15MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13056378
Report Number3008881809-2021-00540
Device Sequence Number17115644
Product Code OUT
UDI-Device Identifier07613327375190
UDI-Public07613327375190
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Model NumberFDS40015
Device Catalogue NumberFDS40015
Device Lot Number22451793
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2021
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date03/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age44 YR
Patient SexFemale
Patient Weight60 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-