Model Number KD-612L
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pulmonary Emphysema (1832); Pleural Effusion (2010); Perforation of Esophagus (2399)
Type of Reportable Event Serious Injury
Event or Problem Description
Olympus medical systems corp.(omsc) received a literature titled "a case of conservative treatment for a large perforation during esophageal esd with suspected partial esophageal strictures".The literature reported the result of a patient with early-stage esophageal cancer who underwent endoscopic submucosal dissection (esd) procedure using olympus dual knife j and it knife nano.In the literature, the following perforation have been reported to have occurred."after marking the periphery of the lesion with a dual knife j, the dissection was started from the mouth side.During the dissection, the endo track was used for traction, but suddenly a huge loss of muscle layer and fibrous structures were observed.The perforation was sutured using clips and indwelling snares, and the perforation was filled with pga sheet and fibrin glue.A ct scan after the procedure showed mediastinal emphysema around the perforation and across the diaphragm.On the next day, the mediastinal emphysema worsened and subcutaneous emphysema of the anterior chest and right pleural effusion appeared, and the inflammatory response increased with wbc 13800 and crp 17.Lmg/dl.After discussion with the surgeon, it was decided to continue conservative treatment because the patient's general condition was stable.Fasting, central venous feeding, and antibiotics were administered, and the inflammatory response gradually decreased.However, gastrography under a fine endoscope performed on the seventh day of the disease showed that the perforation remained and contrast medium leaked out of the esophageal wall from the same site.Since surgical closure of the foramen magnum was invasive and high-risk, fasting was continued, and fibrin glue was reinjected into the foramen magnum on the 12th day.On the 16th day, the foramen magnum was filled with white substance and no leakage of contrast medium was observed.The patient resumed eating on the 17th day and was discharged home on the 26th day." based on the available information, specific information on the subject device and the patients were not provided.There is no description of the device's malfunction.However, omsc assumes that "perforation" might be related to the subject device, and the subject device might be caused or contributed to a death or serious injury.Therefore, omsc determined that the "perforation" was adverse event to submit.Omsc will submit a medical device report (mdr) depending on the event.
 
Additional Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.Since the serial number is unknown, the device history record could not be reviewed.However, omsc has only shipped devices that passed the inspection.In the literature, there is no description of the device's malfunction.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE KD-612 1
Common Device Name
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13057646
Report Number8010047-2021-16575
Device Sequence Number19247873
Product Code KNS
UDI-Device Identifier04953170325892
UDI-Public04953170325892
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K092309
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberKD-612L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2021
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date12/21/2021
Supplement Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexMale
-
-