| Model Number |
8670001 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider via a manufacturer representative regarding a patient with spinal therapy.It was reported that the guide wire got stuck in the hollow part of the screw and when removing it, the screw was removed together.In the s2 sacral foramen part, screw insertion mark was formed.In an additional surgery the surgical procedure was changed from s2ai to iliac screw during the same day.There was no fragment left in the patient.There were no further complications reported regarding the event.Pre-operative diagnosis for this procedure is adult spinal deformity.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis : part # 8670001, lot # kh20m190 visual inspection confirmed the guidewire has been bent and jammed into the cannulated bone screw.This type of damage is consistent with bend stress overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider via a manufacturer representative regarding a patient with pre-op diagnosis-adult spinal deformity involved in the posterior spinal correction fusion procedure.Levels implanted- t11-il.It was reported that the guide wire got stuck in the hollow part of the screw and when removing it, the screw was removed together.In the s2 sacral foramen part, screw insertion mark was formed.In an additional surgery the surgical procedure was changed from s2ai to iliac screw during the same day.There was no fragment left in the patient.There was delay in overall procedure time for less than 60 minutes.Patient was not hospitalized prolong as a result of this event.Product used correctly according to the directions given in the ifu/labeling.There were no further complications reported regarding the event.
|
| |
|
Search Alerts/Recalls
|