| Model Number |
6209-S1 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the introducer was returned and analyzed.Analysis indicated there was a mechanical issue with the dilator of the lead introducer.The sheath of the lead introducer was torn.The sheath of the lead introducer was kinked/buckled.Visual analysis of the introducer indicated damage during use.The analyst noted the introducer returned with the dilator still inserted in it.Visual inspection found the dilator tip melted and its distal end bent.The dilator tip found melted, and this stated in the reported of ¿plastic that had been exposed to heat for too long and becomes stiff¿.The introducer sheath was tear due to kinked/buckled near to the wings.These finding indicated force was applied during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the rigidity/stiffness of the introducer was noted on the tip of the dilator.The introducer was not required. no patient complications have been reported as a result of this event.
|
| |
|