• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX RX25 OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX RX25 OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX-RX25RW
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Event Description
The user facility reported that the capiox device was used pre-treatment.The luer connector on the red cock side dislodged from the sampling system.The oxygenator was changed out.The patient was not harmed.The procedure outcome was not reported.
 
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation- clinical engineer.The actual sample was received for evaluation.Visual inspection confirmed that the lock adapter had come off the male connector of the sampling system.Magnifying inspection of the actual sample did not find any visible anomaly, including deformity, in the male connector or in the lock adapter.The outer diameter of the rib of the male connector and the inner diameter of the lock adapter were measured.Compared to a current product sample, no difference was observed in the measured values.The surface of the male connector was subjected to elemental analysis by sem-edx (scanning electron microscope / energy dispersive x-ray spectroscopy).The result showed the presence of si, which is likely to be derived from the silicone applied to the switch cocks of the three-way stopcock to improve the lubricity of them in the manufacturing process.Simulation test: silicone was applied to a male connector of a factory-retained sampling system, a female connector was attached to it, and then a lock adapter was tightened up.As a result, the lock adapter came off the male connector.Product structure: the sampling system is designed so that the internal step of the lock adapter is hooked on the male connector's ribs to prevent loosening when the female luer is connected.Therefore, if the internal step completely overcomes the ribs for some reason, the mating may come off.A review of the device history record and the product-release judgement record of the involved product code/lot# combination was conducted with no findings.Ifu states: do not use if the package or device is damaged (e.G.Cracked) or any of the port caps are off.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It was likely that the silicon, which is applied to the switch cocks of the sampling system to improve the lubricity, was transferred to the male connector part due to some factors; therefore, it was likely that the lock adapter came off when it was re-tightened.However, from the state of the actual sample, the exact cause of the reported event cannot be definitively determined based on the available information.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX25 OXYGENATOR
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key13058968
MDR Text Key289755680
Report Number9681834-2021-00244
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040210
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2024
Device Catalogue NumberCX-RX25RW
Device Lot Number210802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2021
Initial Date FDA Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-