| Model Number |
G5A-80A |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report was duplicated and evaluation of the device found that program firmware installed had become corrupted.The program firmware was re-installed to resolve the report.It could not be firmly established what caused the firmware to become corrupted.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|