It was reported that during preparation of an emboshield nav6, during loading of the filtration element into the delivery catheter (dc) pod, it was noted that there was bunching/wrinkling observed on the dc pod.The filter was able to be loaded when it was also noted that the tip of the dc pod was separated.The device was not used and there was no patient involvement.Another nav6 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
Visual analysis was performed on the returned device.The reported material separation and damage to the dc pod was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints.The investigation was unable to determine a cause for the reported damage to the delivery catheter pod.It may be possible that the damage occurred during preparation and/or removal from the loading tray; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|