| Model Number |
30104 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At the time of this investigation, the device history record could not be verified as a lot number was not provided.As no samples were returned for evaluation, a root cause could not be determined.We will continue to monitor complaint trends and utilize the information as part of continuous improvement.
|
| |
|
Event or Problem Description
|
|
Customer reported that they received cardinal health hot packs that burst during transport while in the cardinal health totes.There was no injury reported and no one at the facility was exposed.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental report is being filed since samples from lot number v1l109 were returned for investigation.The lot was found to have been manufactured and released to predetermined specifications.No anomalies were found during review of the records.Ten samples were received for evaluation.Eight of the samples were not activated or leaking.One sample was pre-activated and one was pre-activated and torn.The root cause could be a torn pouch due to the bubble puncturing the pouch; it cannot be determined when the bubble punctured the pouch.Increased handling during shipping and processing are elements that could prompt pre-activation.We will continue to monitor complaint trends and utilize the information as part of continuous improvement.
|
| |
|