Model Number 11007642
Medical Device Problem Code Unintended Application Program Shut Down (4032)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported event.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.Internal id # (b)(4).
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the sensis vibe hemo system.During an interventional procedure, the user reported that the device was totally down following system reboot.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.The reported event occurred in (b)(6).
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.According to the available information, the system reminded the user of the 7-day reboot via a pop up window.After the system reboot, the system did not boot accordingly and the procedure was performed using an alternative system.The investigation could not clarify the root cause in detail as the error has been resolved by the customer by a repeated reboot.Furthermore, no log files were provided for a detailed analysis.Per our system experts; a bios setting problem could potentially lead to the sporadic error state.A solution is currently in development.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIS VIBE HEMO
Common Device Name
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
mc 65-1a
malvern, PA 19355
6104486461
MDR Report Key13061197
Report Number3004977335-2021-09878
Device Sequence Number1608639
Product Code DQK
UDI-Device Identifier04056869010199
UDI-Public04056869010199
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K150493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11007642
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2021
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date12/22/2021
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-