| Model Number |
11007642 |
| Medical Device Problem Code |
Unintended Application Program Shut Down (4032)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens is conducting a thorough investigation of the reported event.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.Internal id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction occurred while operating the sensis vibe hemo system.During an interventional procedure, the user reported that the device was totally down following system reboot.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.The reported event occurred in (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.According to the available information, the system reminded the user of the 7-day reboot via a pop up window.After the system reboot, the system did not boot accordingly and the procedure was performed using an alternative system.The investigation could not clarify the root cause in detail as the error has been resolved by the customer by a repeated reboot.Furthermore, no log files were provided for a detailed analysis.Per our system experts; a bios setting problem could potentially lead to the sporadic error state.A solution is currently in development.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
|
| |
|
Search Alerts/Recalls
|