• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HUBER PLUS 20G X 1" NEEDLELESS Y SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HUBER PLUS 20G X 1" NEEDLELESS Y SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refs1473 showed three other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that three needles exhibited leakage from y site connector.No other information was provided.This report addresses the second device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a leak at the y-site hub is confirmed but the exact cause is unknown.Two huber plus 20 ga x 1 in infusion set devices were returned in open packaging.Both trays exhibited deformation which appeared to be caused by heat damage.The flanges were curled and warped.The tyvek lid was returned peeled; however, evidence of a complete seal was observed.An inspection of the infusion sets showed the y-site hubs to contain an oval shaped deformation.The infusion sets were flushed with water and leaks were noted at the y-site hubs.Microscopic observation of the y-site revealed a gap between the blue valve and white housing due to the deformation.The exposure to heat appears to have affected the warping of the tray packaging and infusion sets.This type of damage can occur from prolonged exposure to excessive heat, and the damage was most likely associated with shipping or environmental conditions outside bd control.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported that three needles exhibited leakage from y site connector.No other information was provided.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUBER PLUS 20G X 1" NEEDLELESS Y SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13061270
Report Number3006260740-2021-05402
Device Sequence Number11139561
Product Code FPA
UDI-Device Identifier00801741065569
UDI-Public(01)00801741065569
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K993848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number012001NY
Device Lot NumberREFS1473
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date12/22/2021
Supplement Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-