| Catalog Number |
2000-2450 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2021-00478 through 3012447612-2021-00480.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, three multi-axial screw heads jammed so they lost their polyaxial motion.New screws were used to complete the case without patient impacts.This is report one of three for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, three multi-axial screw heads jammed so they lost their polyaxial motion.New screws were used to complete the case without patient impacts.This is report one of three for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: functional inspection revealed that each screw was unable to maintain polyaxial movement without significant jamming/sticking.Root cause: root cause was unable to be determined.This event could possibly be attributed to unknown events during handling or use.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|