| Model Number |
393055 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) 2021, eri alert was received via hm.The date before (10 dec.2021) had 45 percent battery remaining, but todays battery showed 0 percent remaining.(b)(6) 2021, the device was explanted and replaced.Please note this icd is affected by a field safety corrective action, bio-lqc, initiated in (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.As a first step in the analysis, the device was interrogated.The interrogation could be properly performed, and it revealed the battery status eri.Further analysis of the icd memory content showed a normal current consumption.However, the amount of charge taken from the battery was verified and the battery condition was not as expected.In order to obtain further insights, the icd was opened, and the inner assembly was inspected.The visual inspection showed no anomalies.The overall current consumption of the electronic module was verified by direct measurement and proved to be normal and as expected.There was no indication of a malfunction of the electronic module.Battery voltage measurement confirmed a depleted battery that can be attributed to an increased internal self-depletion within the battery as a result of lithium plating.This icd is part of the population affected by the field safety notification regarding potential premature battery depletion due to lithium plating from march 2021.
|
| |
|
Search Alerts/Recalls
|