| Model Number |
TN, LH780 ANALYTICAL STATION |
| Medical Device Problem Codes |
Fire (1245); Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported while turning on the main switch for the compressor following a power surge from a thunderstorm; a burst inside the compressor was heard with a small flame and smoke were seen coming from the sound absorbing padding of the compressor in the lh780 hematology instrument.The power supply was immediately disconnected.The customer then doused instrument with water.A fire extinguisher was not used.The fire was contained within the instrument.The fire department was not called no medical attention needed, there was no death, injury, or change to patient or customer due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was at the customer site and found that the sound absorbing padding of the compressor unit was burnt.The instrument was not repaired and was scrapped.The customer is replacing the instrument with a newer model instrument.Per staff product compliance engineer, a definitive hazard cannot be identified without a physical evaluation.However, since it was reported the event was contained in the compressor cabinet; it was surmised that no fire hazard existed as a result of this event.(b)(4).
|
| |
|