| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Codes |
Output Problem (3005); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user has been reimplanted with a new device on (b)(6) 2021.As per explant report form the explantation was due to loss of function.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The recipient has been reimplanted with a new device on (b)(6)2021.As per explant report form the explantation was due to loss of function.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.The problems given in the recipient report appear to match the damage found.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
|
| |
|