• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID INT TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-65
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device : a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and tested the unit by doing the helium pressure test, but the unit was still down 39 psi.The gasket was replaced, tightened the helium cylinder and checked the docking o-ring, but the issue could not be resolved.A supplemental report will be submitted when additional information is provided.The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
 
Event Description
It was reported that during an inspection the cardiosave intra-aortic balloon pump (iabp) was not passing the helium pressure test.There was no patient involvement, and no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Additional information: e1(event site postal code: (b)(6).Event site state:(b)(6)).Additional point of contact: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID INT TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13076891
MDR Text Key282713656
Report Number2249723-2021-02943
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-65
Device Catalogue Number0998-00-0800-65
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/23/2021
Supplement Dates Manufacturer Received02/20/2024
Supplement Dates FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-