Model Number 0998-00-0800-65 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device : a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and tested the unit by doing the helium pressure test, but the unit was still down 39 psi.The gasket was replaced, tightened the helium cylinder and checked the docking o-ring, but the issue could not be resolved.A supplemental report will be submitted when additional information is provided.The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
|
|
Event Description
|
It was reported that during an inspection the cardiosave intra-aortic balloon pump (iabp) was not passing the helium pressure test.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Additional information: e1(event site postal code: (b)(6).Event site state:(b)(6)).Additional point of contact: (b)(6).
|
|