| Model Number |
1219-32-054 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/06/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision of tha completed.Acetabular cup and poly liner was dissociated from one another.Liner had significant wear and tear.Original components were replaced with 56mm cup, a marathon polyethylene +4 neural 36/56 liner and 36mm +1.5 femoral head.Doi: unknown, dor: unknown, unknown side.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.Reviewing the pictures provided on the attachments we can confirm the reported event.The liner shows disassociation evidence.It was not possible however to identify any product error as the root cause for the event.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that there was no surgical delay reported.The affected side involved in this event was the right hip.
|
| |
|