| Model Number |
86200 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an ablation to treat left atrial flutter a ultra ice plus catheter and rhythmia mapping system were selected for use.It was reported that poor ultrasound imaging/noise was observed during the procedure.The catheter was replaced, however, the transeptal puncture was unsuccessful.The procedure was aborted.Further troubleshooting indicated that the cause of the noise was the rhythmia signal station.
|
| |
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|