| Model Number |
OPTI0930CSFMX |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Balt usa reference #: (b)(4).An evaluation of the actual complaint sample could not be performed as the device was unavailable for return.Based on the provided information, root cause could not be definitively determined.Lack of device return prevented deeper evaluation of the reported issue.Review of the lot history records did not reveal any in-process or lot-specific issue that could account for the observation.No additional complaints against lot number f210200495 has been made for the same issue.Comprehensive analysis of this failure mode has remained subject to monitoring for any unacceptable increase in trend.
|
| |
|
Event or Problem Description
|
|
It was reported that: "premature detachment of the coil during coil positioning into the embolization site.Microcatheter used: excelsior 17xt straight.Coil retrieved with a lasso system." follow up (b)(6) 2021.Additional information received from issuer:" was there torturous anatomy involved? no." follow up (b)(6) 2021: additional information received from issuer: "where did the implant prematurely detach? did it occur inside of the microcatheter? absolutely.".
|
| |
|