ENCORE MEDICAL L.P. EXPRT KNEE; INSERT,REVISION EXPRT, SZ A X 14.5MM, E-PLUS Back to Search Results
Model Number 315-0A-714
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery: patient was unstable.
 
Additional Manufacturer Narrative
The reason for this revision surgery was instability.The previous surgery and the surgery detailed in this event occurred 4 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this compliant was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, in adequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT KNEE
Common Device Name
INSERT,REVISION EXPRT, SZ A X 14.5MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, TX 78758-5445
MDR Report Key13079046
Report Number1644408-2021-01474
Device Sequence Number4153799
Product Code JWH
UDI-Device Identifier00888912113915
UDI-Public(01)00888912113915
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/09/2022
Device Model Number315-0A-714
Device Catalogue Number315-0A-714
Device Lot Number004R1916
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2021
Initial Report FDA Received Date12/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
-
-