Model Number 45007 |
Device Problem
Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that a loss of aspiration occurred.The target lesion was located in the superficial femoral artery (sfa).A 2.1mm jetstream xc atherectomy catheter was selected for an atherectomy procedure.A lot of thrombotic material was present.The catheter stopped sucking and was removed from the patient.The procedure was successfully completed with another 2.1mm jetstream catheter.There were no patient complications and the patient's status was fine.
|
|
Manufacturer Narrative
|
Date of event: no date provided, used the first date in the month of the aware date.
|
|
Event Description
|
It was reported that a loss of aspiration occurred.The target lesion was located in the superficial femoral artery (sfa).A 2.1mm jetstream xc atherectomy catheter was selected for an atherectomy procedure.A lot of thrombotic material was present.The catheter stopped sucking and was removed from the patient.The procedure was successfully completed with another 2.1mm jetstream catheter.There were no patient complications and the patient's status was fine.
|
|
Manufacturer Narrative
|
B3.Date of event: no date provided, used the first date in the month of the aware date.Device evaluated by mfr: returned product consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed no damage.Functional analysis was done by completing the setup procedure per the instructions for use.Aspiration testing of the device was done per the test procedure.The device is tested by using a 100ml beaker of water.The devices tip is submerged in a beaker of fluid and the device is run for a period of 1 minute.The final milliliters of fluid that the beaker shows after the 1minute run time is subtracted from the starting milliliters and the final number is the total that was aspirated (100ml-2ml=98ml).The minimum amount of fluid that is aspirated per the test procedure specification sheet is 30ml per minute and the maximum is 56ml per minute.Test results showed that this device did not perform as designed per the test procedure specification sheet withdrawing 2ml of fluid in the 1minute time frame.Inspection of the remainder of the device revealed no damage or irregularities.
|
|
Search Alerts/Recalls
|