| Model Number |
EN28162400 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to available information, this device required revision due to a hole in a cylinder.The reservoir was retained.No other adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump, cylinders 1 and 2, and detached inlet tube with connector were received for evaluation.Abrasion was noted on the longer exhaust tube and inlet tube of the pump.No functional abnormalities were noted with the pump.A separation was noted in the exhaust tube of cylinder 1 at the tube/strain relief junction.This was a site of leakage.The surfaces appeared to be rough and irregular, indicating stress was exerted.A partial separation was noted on the exhaust tube of cylinder 1.This was not a site of leakage.Abrasion was noted on both cylinder exhaust tubes.No functional abnormalities were noted with cylinder 2.Abrasion was noted on the inlet tube.No functional abnormalities were noted with the tube or connector.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.This positioning, in combination with device usage over time, may contribute to sufficient stress(s) to separate the exhaust tubing at the tube/strain relief junction of cylinder 1.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Search Alerts/Recalls
|